Pass-through ovens UF TS are based on a standard heating oven and feature all technological highlights like product specific heating and perfectly adjusted control technology. Thanks to an additional side feedthrough, curing of lead frames and adhesive bonds or tempering of components can be controlled automatically within a running production process.
- ControlCOCKPIT: TwinDISPLAY with two high-resolution TFT colour touchdisplays
- Voltage: 230 V, 50/60 Hz for sizes 160/260
- Voltage: 400 V, 50/60 Hz for sizes 450/750
- Setting temperature range +20 °C to +250 °C
- For 2-sided loading (door front and back)
- All-round heating on all 4 sides for excellent temperature homogeneity
- Admixture of pre-heated fresh air by electronically adjustable air flap
- The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at all measuring points – optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber
- Internal data logger with storage capacity of at least 10 years
- Works calibration certificate at +160 °C
- Supplied with 2 x Stainless steel grids, electropolished (standard)
Warranty is 3 years on parts, 1 year on Labour
TwinDISPLAY:
· ControlCOCKPIT with two high-resolution TFT colour touchdisplays
· Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program time, time zones, daylight-saving time
· Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
· HeatBALANCE function for application specific adjustment of heat output distribution (balance) between the upper and lower heating groups in an adjustment range between -50 % and +50 %
· AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
· Intuitive programming and archiving of ramps and program sequences
· Synchronous visualisation of the created program sequence during programming
· Application-specific repeat functions (loops) can be inserted within a temperature control program in any place
· Simple creation of repeating weekly programs
· ControlCOCKPIT with USB port for uploading programs, reading out protocol logs, activating the User-ID function
· Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
· Ethernet interface for reading out protocol log, uploading programs and for online logging
· Multiple over temperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2) and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual
· MIN / MAX values for over/under temperature alarm
This product has met the following criteria: