General description
Low power usage Easy to wall mount CE approvedNitrox Zero Air Generators employ a catalytic oxidation system to produce a continuous supply of air that is virtually free of hydrocarbons, including methane. After incoming compressed air is filtered to remove oil, water, and particles, it is passed over a heated platinum catalyst that converts methane and oxygen in the air to carbon dioxide and water. If the hydrocarbon content of the input air is less than 10 ppm (as methane), the residual hydrocarbon concentration (as methane) is less than 0.1 ppm. A high efficiency 0.01 micron filter downstream of the catalyst ensures further removal of particles from the product stream.Maintenance filter kits and warranty infomation available upon request. Contact Supelco Technical Service at techservice@sial.com or phone 800-359-3041.
Legal Information
domnick hunter is a registered trademark of Parker Hannifin
This product has met the following criteria: