Memmert Humidity chamber HCP50 56L with 1 stainless steel shelf

Code: CHA4010 N1-1

Applications for humidity chambers range from construction physics to corrosion testing and down to biological research. Ramp programming for temperature and humidity, active humidity control betwe...


read more

Your Price
£7,828.00 EACH
£9,393.60 inc. VAT

3 Year Warranty
Service Centre
Calibration Service

Applications for humidity chambers range from construction physics to corrosion testing and down to biological research. Ramp programming for temperature and humidity, active humidity control between 20 and 95 % rh as well as exact temperature control of up to +90 °C ensure a controlled, physiologically ideal environment for the simulation of real conditions.

  • ControlCOCKPIT: TwinDISPLAY with two high-resolution TFT colour touchdisplays
  • Setting temperature range 18 °C to 90 °C
  • Setting range active humidity control 20 to 95 % rh
  • Capacitive humidity sensor for measuring and displaying the relative humidity
  • Fully insolated stainless steel door and heated inner glass door
  • All-round heating on 6 sides
  • Ethernet interface for reading out protocol log
  • USB port on ControlCOCKPIT for use of USB stick
  • Multiple overtemperature protection/undertemperature alarm
  • Works calibration certificate at 60 °C / 75 % rh
  • Warranty is 3 years on parts, 1 year on Labour

TwinDISPLAY:

  • ControlCOCKPIT with two high-resolution TFT colour touchdisplays
  • Available parameters on the ControlCOCKPIT: Temperature (Celsius or Fahrenheit), fan speed, exhaust air flap position, program time, time zones, daylight-saving time
  • Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error
  • HeatBALANCE function for application specific adjustment of heat output distribution (balance) between the upper and lower heating groups in an adjustment range between -50 % and +50 %
  • AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
  • Intuitive programming and archiving of ramps and program sequences
  • Synchronous visualisation of the created program sequence during programming
  • Application-specific repeat functions (loops) can be inserted within a temperature control program in any place
  • Simple creation of repeating weekly programs
  • ControlCOCKPIT with USB port for uploading programs, reading out protocol logs, activating the User-ID function
  • Displaying of already logged protocol data on the ControlCOCKPIT (max 10,000 values correspond to approx. 1 week)
  • Ethernet interface for reading out protocol log, uploading programs and for online logging
  • Multiple over temperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1 or 2) and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual
  • MIN / MAX values for over/under temperature alarm

CodeModelVolume, LStainless Steel GridsInt Dims*, w x d x h, mmExt Dims**, w x d x h, mm
CHA4010HCP50561400 x 330 x 425559 x 521 x 795
CHA4012HCP1051072560 x 400 x 480719 x 591 x 850
CHA4014HCP1501562560 x 400 x 700719 x 591 x 1070
CHA4016HCP2402412600 x 500 x 810759 x 691 x 1180
* Internal dimensions: depth less 35mm for fan
** External dimensions: depth without door handle, depth of handle 56mm
Code
Description
Unit Size
List Price
Qty
CHA4012
Unit:EACH
List Price: £8,896.00
Source:List Price
ADD
CHA4014
Unit:EACH
List Price: £12,186.00
Source:List Price
ADD
CHA4016
Unit:EACH
List Price: £13,471.00
Source:List Price
ADD
Int Dims, w x d x h, mm400 x 330 x 425
Temp Range, °C18 to 90
Weight, kg55